Subject: Re: New Blue system testing |
From: Casey Smith |
Date: 4/24/2015 3:35 PM |
To: yarbrough |
CC: feinholz, Carol, Stephanie , Kenneth Voss , Michael Kehoe |
Happy Aloha Friday,
I did a course alignment of the returned Moby this afternoon,
with a modified slit mount to decrease the distance between the
fiber face and the slit.
The spatial sharpness of the fiber channels is much better.
Please see the attached images.
I will be working on this next week, trying to better center the
fiber image on the detector and to finalize the other parameters,
and will have more data for you then. I just wanted to get this out
before the weekend.
cheers,
casey
-- Casey Smith, Research Scientist Resonon, Inc. 619 N. Church Ave Ste 3 Bozeman, MT 59715 (406) 586-3356 x 114 smith@resonon.com www.resonon.com